ResMed CPAP Error Codes and Troubleshooting

  • Writer: Stubborn Shreya
    Stubborn Shreya
  • Jun 12
  • 10 min read

Updated: Jul 3


Introduction: When Your ResMed Machine Throws an Error—Here’s What It Means

Whether you’re using a trusted AirSense 10, the sleek new AirSense 11, or a robust Lumis VPAP, ResMed machines are designed for reliability and comfort. But like any advanced device, they sometimes flash an error code, make an unusual sound, or simply stop working mid-therapy. These alerts can feel intimidating—especially if they appear in the middle of the night. In this guide, we’ll break down ResMed error codes across all major models, from 001 to E-24, so you know exactly what they mean, how serious they are, and what steps you can take to fix them. Whether it's a humidifier fault, pressure issue, or firmware error, our goal is to help you resolve it quickly—and sleep better. CPAP मशीन एरर कोड को समझना ज़रूरी है ताकि आप सही समय पर इलाज और मशीन रिपेयर करवा सकें।


Which ResMed Devices Are Covered in This Guide

This guide focuses on the most widely used and searched-for ResMed CPAP and BiPAP models based on real-world user queries and verified device manuals. Whether you're using the AirSense 10 AutoSet, the AirSense 11 AutoSet, a Lumis 100 or 150 VPAP ST, or even an older ResMed S9, you'll find the exact error codes and their meanings listed here. We’ve also included a brief section on AirMini and cross-model errors—covering recurring alerts like Error 001, motor stall, and humidifier faults that tend to appear regardless of the machine model. This structured approach helps you quickly locate the exact fix based on the machine you’re using at home. Which CPAP errors do users in Delhi NCR most commonly report?


ResMed AirSense 10 AutoSet / Elite – Error Codes and Fixes

The ResMed AirSense 10 AutoSet and Elite are some of the most trusted machines used for treating obstructive sleep apnea. ResMed मशीन की सबसे आम समस्याएं क्या हैं? चलिए जानते हैं इनके बारे में हिंदी में। Can you adjust the ResMed AirSense 10 pressure setting on your own in Delhi? But over time, they may display error codes or alerts tied to blowers, sensors, SD cards, or humidifiers. Here’s a breakdown of the most common problems users report and what each message actually means:


Error / Message

What It Means

Fix / Action

Error 001

Hardware fault or internal error

Requires service; may need PCA replacement

Error 006 / 007 / 008

Blower motor stalled or failed

Replace blower motor or mainboard

Error 010 / 011 / 012

Flow sensor stuck or not transmitting data

Replace flow sensor or main PCB

Error 013

Pressure sensor malfunction or reading issue

Recalibrate or replace main PCB

Error 017 / 018

Software or memory-related faults

Service required; mainboard may need replacement

Power connector not fully inserted

Adapter is loose or not seated

Reseat the power cord securely into the port

Motor life exceeded

Blower motor has hit 21,900-hour threshold

Contact service center to replace motor

Water leaking from chamber

Chamber cracked or misaligned

Replace chamber; ensure it's seated properly

Blocked inlet detected

Filter clogged or air path blocked

Clean or replace air filter, clear intake

High leak detected

Air escaping from mask or humidifier

Reseat mask, check seal, tighten tubing

SD card error / corrupted

SD card not detected or damaged

Reinsert, reformat, or replace the SD card

Device temperature too high

Overheating due to environment or motor stress

Move to a ventilated area; power off if it persists

Software upgrade in progress

Firmware update running

Allow 10+ minutes; don’t unplug during update


These alerts may show on-screen with icons, flashing messages, or via a red light. Treating them early—especially motor or SD card errors—prevents escalation into full service-required conditions. अगर आपकी CPAP मशीन बार-बार E-83 या E-89 एरर दिखा रही है, तो तुरंत सर्विस कराएं। Can you fix the E-89 error on ResMed AirSense 10 yourself in Delhi or Gurgaon, or do you need a professional?


🛠️ Facing issues with your ResMed CPAP machine in Noida or NCR? We offer doorstep service, rental & sales.


📍 Visit Our Noida Store

📞 Call Immediately, 📱 WhatsApp Now: +91 9876978488


ResMed AirSense 11 Error Codes & Fixes 

While the ResMed AirSense 11 introduced a modern touchscreen and updated algorithms, it still shares many core error types with the AirSense 10. Users have reported numeric codes ranging from Error 1 to Error 24, typically linked to blower faults, pressure sensor errors, flow sensor failures, or firmware glitches. If your ResMed CPAP machine isn’t working in Delhi, can you reset it yourself? The newer interface makes alerts more visible, but the causes and fixes remain rooted in hardware and calibration issues. humidifier सही से जुड़ा है या नहीं, यह भी एक बड़ी वजह हो सकती है — इसे दोबारा जांचें।


Error Code / Message

Cause

Recommended Fix

Error 1, 6, 7, 8

Blower motor faults or stalling

Replace motor via authorized service center

Error 3

Pressure sensor calibration failure

Try recalibration; may require PCA replacement

Error 4, 5

Undefined software or circuit error

Contact ResMed support; device likely needs evaluation

Error 9

Mainboard or control failure

Service center replacement of PCA

Error 10–13

Flow sensor failure or signal loss

Requires flow sensor and/or mainboard replacement

Error 14, 15, 17, 18

Software or PCB failure (data/logic layer)

Device needs diagnostics and possible board replacement

Error 21, 23, 24

Severe hardware or calibration fault

Remove SD card, restart; escalate to service if repeated

System Fault: Error 4 / X

General hardware failure

Let device cool, check airflow, contact service if unresolved

Motor life exceeded

Device reached expected lifespan (~21,900 hrs)

Replace blower motor and recalibrate

Water leak / humidifier alert

Faulty chamber or improper seating

Replace or reposition water tank securely

High Leak Detected

Loose mask or tube connection

Inspect mask fit, humidifier lid, and hose attachments

Screen not waking up

Power glitch or app-display sync delay

Tap screen, verify adapter; use MyAir if issue continues

Device not powering on

Bad outlet or adapter

Swap outlets, check green light on brick, replace if needed

With AirSense 11, some alerts may show as pop-ups while others appear during startup. In either case, addressing airflow blockages and checking the power supply often fixes basic faults. For more persistent issues, especially numbered codes, a technician will likely need to replace key components. Where can you find ResMed CPAP repair service in Noida, Delhi, NCR? Find out from Healthy Jeena Sikho.


पंजाब और चंडीगढ़ के मरीजों के लिए – Healthy Jeena Sikho से CPAP मशीन किराये पर लें या सर्विस कराएं।

📞 अभी कॉल करें या 📱 WhatsApp करें: +91 9876978488


ResMed Lumis 100/150 VPAP ST – Error Codes & Fixes 

The ResMed Lumis 100 and 150 VPAP ST are designed for patients with serious respiratory conditions requiring non-invasive ventilation. These machines use a more technical error reporting system, which includes both numeric codes (e.g. Error 206, 409, 420) and priority-based alarms. Many errors stem from issues with circuit compliance, oxygen calibration, or internal hardware faults.


Error Code / Message

Cause

Recommended Fix

206, 404–406, 415, 426

General hardware failure

Contact a ResMed service center for full diagnosis and repair

Error 409 (Learn Circuit Failed)

Leak or misassembled circuit

Check for leaks, refit the circuit, and repeat the Learn Circuit test

Error 420 (High Compliance)

Circuit flow resistance out of range

Block the circuit temporarily and re-run Learn Circuit

Error 303 (Oxygen sensor calibration failure)

Oxygen added during Learn Circuit

Ensure no supplemental oxygen during test; rerun Learn Circuit

Error 004 / 005 / 006

Internal software, blower, or pressure sensor faults

Restart device; if repeated, service required

Error 008

Humidifier or heating plate issue

Remove and reseat humidifier; replace if unresolved

Blocked Tube Detected

Tube is occluded or kinked

Straighten or replace tubing

Power Supply Error

Incorrect adapter or unstable voltage

Use official ResMed adapter only

SD Card Not Detected

Card missing, corrupted, or incorrectly inserted

Reinsert or replace card; format if needed

Water Tub Leak / Humidifier Error

Faulty seating or cracked chamber

Reseat tub and inspect; replace if leak persists

Low Minute Ventilation

Patient breathing below preset threshold

Check patient condition and settings

Apnea Alarm

Breathing paused longer than limit

Check mask seal and patient response

🔔 Alarm Priority Levels:

  • Red Light + Loud Alarm: Critical—check patient + device immediately

  • Yellow Light + Medium Beep: Moderate issue—check mask/circuit soon

  • Yellow Light + Soft Beep: Low concern—inspect SD card, tubing, or filter

The Lumis series is used in more complex settings, so early response to codes like 409, 420, or 303 can prevent therapy breakdown or loss of circuit calibration. Always re-run the Learn Circuit test after correcting tubing or humidifier faults. Where to contact for buying or renting ResMed CPAP in Delhi, Noida, Chandigarh, or Ludhiana?


ResMed S9 Series – Error Codes & Fixes 

Although the ResMed S9 Series is an older generation of CPAP machines, it’s still widely used in home therapy setups across India. Getting an alarm or red light on your CPAP machine in Delhi? What does it mean? Because of its age, many users encounter hardware degradation errors, humidifier faults, and SD card issues after years of consistent use. Below is a list of the most common ResMed S9 errors and what to do if they appear:


Error / Message

Cause

Recommended Fix

Motor Life Exceeded

Blower motor has reached its 21,900-hour design life

Replace motor and recalibrate via authorized service center

Motor Fault

Blower malfunction or failure

Replace motor at ResMed-certified repair center

Humidifier Fault

Component failure or water damage

Power off, dry thoroughly, and replace humidifier if needed

Invalid PSU / Power Supply Error

Adapter not recognized or incompatible

Use genuine ResMed 30W or 90W power supply; insert plug fully

High Temperature Fault

Device overheating from blocked airflow or external heat

Let cool down, relocate to ventilated area, reconnect power

Air Filter Blocked

Dirty or clogged filter

Remove and replace filter; restart device

Air Tubing Blocked

Kinked or occluded hose

Check for bends or obstructions; reconnect securely

High Leak Warning

Poor mask seal or loose connections

Adjust mask, replace old tubing, check humidifier seal

Water in Tubing

Humidity setting too high for ambient room

Lower humidifier setting and drain tube

Check Power Supply Connection

Plug not inserted correctly or covered

Reinsert DC plug, keep supply uncovered and dry

Card Error / Not Detected

SD card missing, corrupted, or removed incorrectly

Reinsert card, wait for home screen; replace if needed

Calibration Data Corrupted

Board memory failure or software bug

Restart device; if unresolved, PCA board replacement needed

Many S9 users report these issues surfacing around the 4–5 year mark. To extend the device’s life, regular filter changes, proper storage, and using original accessories are key. And always handle SD cards carefully—removing them mid-write is a leading cause of boot errors.


दिल्ली, नोएडा, गुड़गांव, लुधियाना और चंडीगढ़ में हम ResMed CPAP मशीन की रिपेयर सर्विस प्रदान करते हैं। Get ResMed CPAP machines on rent or purchase at best prices in Delhi NCR & Punjab. Doorstep delivery available!


📞 Call Now or📱 WhatsApp Now: +91 9876978488


Frequently Asked Questions (FAQ)


1. What do the most common ResMed CPAP/VPAP error codes mean?ResMed machines display error codes and messages for issues like motor faults, sensor errors, power supply problems, leaks, and humidifier failures. Each code or alert points to a specific problem—such as “Motor life exceeded,” “Blocked tube detected,” or “System fault, refer to user guide.” 


2. How do I fix a “Blocked tube detected” or “Air tubing is blocked” error?

  • Check your tubing for kinks, blockages, or water buildup.

  • Remove and reconnect the tubing to the device.

  • Replace tubing if damaged or persistently blocked


3. What should I do if my ResMed device shows “System fault” or an error code like 004, 005, or 006?

  • Power off the device and unplug it for at least 1 hour.

  • Move it to a cool, dry area and restart.

  • If the error persists, contact your provider or a ResMed Authorized Service Center, as these usually indicate hardware or sensor faults requiring professional repair.


4. Why does my machine say “Motor life exceeded, contact care provider”?This means the blower motor has reached its design life (about 21,900 hours or 5 years of nightly use). The device needs a new blower motor, which must be replaced and recalibrated by an authorized service center


5. What causes “High leak detected” or “Check mask fit” messages?

  • Mask may be loose or incorrectly fitted.

  • Tubing connections may not be secure.

  • Humidifier chamber or seals may be leaking.

  • Adjust your mask, check all connections, and replace worn components as needed


6. How do I resolve SD card errors or “Card error, please remove SD card and contact service provider”?

  • Remove and reinsert the SD card.

  • If the error persists, try a new SD card.

  • Only remove the SD card when the device is off or not writing data.


7. My device won’t power on. What should I check?

  • Ensure the power cord is fully inserted and the outlet is working.

  • Use only a genuine ResMed power supply.

  • Try a different outlet or power supply if available.

  • If still unresponsive, contact your provider


8. Can I clean and reuse my ResMed filter?No, ResMed filters are not designed to be cleaned—replace them regularly as recommended in your user manual.


9. Why is water leaking from my CPAP/VPAP machine?

  • The water chamber may be cracked or not seated properly.

  • Always use a genuine ResMed water chamber and ensure it is installed correctly.


10. What should I do if my device overheats or shows a high temperature fault?

  • Move the device to a well-ventilated area.

  • Allow it to cool before restarting.

  • Make sure the power supply is not covered or obstructed.


11. What is the ramp feature and how does it help?The ramp feature starts therapy at a lower pressure, making it easier to fall asleep. The device gradually increases pressure to your prescribed level over a set time.


12. How do I find my device’s serial number or device number?Both numbers are located on the label at the back of your device.


Need Help with Your CPAP or BiPAP Machine?

At Healthy Jeena Sikho, we specialize in providing complete respiratory care solutions — including ResMed CPAP and BiPAP machines — for rental and sale. Whether you’re facing machine errors, need expert setup, or looking for affordable equipment on rent, our experienced team is here to help. We proudly serve across Delhi, Noida, Gurgaon, Chandigarh, Ludhiana, and Jalandhar, with same-day delivery and on-site support.


📞 Call us today at +91 9876978488 for expert advice,📱 Or message us directly on WhatsApp for quick assistance.


CPAP मशीन को किराये पर लेना हो या रिपेयर कराना हो – हमें WhatsApp करें: +91 9876978488


Conclusion

From the feature-packed AirSense 11 to the legacy workhorse ResMed S9, every ResMed CPAP or VPAP device is built with safety checks designed to protect you. But when error codes like 001, 420, or motor life exceeded flash on the screen, it's natural to feel alarmed—especially during the middle of the night. This guide helps you decode these messages model by model, giving you the clarity to act fast and get back to therapy. Whether it’s reseating your SD card, replacing a humidifier chamber, or knowing when it’s time to call a service technician, your response can make the difference between a short pause and a prolonged therapy break. Bookmark this page, keep your model's error codes handy, and rest easy knowing you’re prepared for any alert that comes your way.


To know more about BiPAP Machines or CPAP Machines, - Check out our related Articles


 
 
 

4 Comments

Rated 0 out of 5 stars.
No ratings yet

Add a rating
Guest
Jul 03
Rated 4 out of 5 stars.

Can you include a section on how often these error codes occur and whether they indicate serious issues?

Like

Guest
Jul 03
Rated 5 out of 5 stars.

Very helpful! I was confused by the E-28 error on my ResMed AirSense 10 — your explanation made it easy to understand and fix. Thanks!

Like

Stubborn Shreya
Stubborn Shreya
Jul 03
Rated 5 out of 5 stars.

Great breakdown of common CPAP errors! Would love to see a printable PDF version of this guide for quick reference

Like

nice
Jun 13
Rated 5 out of 5 stars.

nice

Like
bottom of page